Background As minimally invasive spine surgery becomes more widespread, concerns regarding radiation exposure to surgeons and patients alike have become a growing concern. Robotic guidance has been developed as a way to increase the accuracy of instrumentation while decreasing radiation burden.Methods A retrospective analysis of a large, multi-centre, prospective study comparing robotic-guided (RG) to fluoroscopic-guided (FG) (Multi-centre, Partially Randomized, Controlled Trial of MIS Robotic vs. Freehand in Short Adult Degenerative Spinal Fusion Surgeries) was performed to evaluate for differences in radiation exposure between study groups.Results RG was associated with 78.3% (p < 0.001) and 79.8% (p < 0.001) reduction in total and per screw fluoroscopy times, respectively, as compared to FG. RG was also associated with a 50.8% (p < 0.001) reduction in total operative fluoroscopy time.Conclusions RG was associated with significantly lower fluoroscopy times compared to FG. This suggests that utilization of robotic navigation systems may result in decreased operative radiation exposure, which is a growing concern for surgeons performing minimally invasive spine surgery.