A MULTICENTER EVALUATION OF THE A-AND-D TM-2420 AMBULATORY BLOOD-PRESSURE RECORDER

WB WHITE,TG PICKERING, J MORGANROTH,GD JAMES,EJ MCCABE, O MOUCHA, H HUNTER

AMERICAN JOURNAL OF HYPERTENSION(1991)

引用 59|浏览3
暂无评分
摘要
The A&D TM-2420 (A&D Engineering, Milpitas, CA) is an automatic, portable, noninvasive blood pressure (BP) recorder which uses a dual microphone system for the detection of Korotkoff sounds. Its accuracy and clinical performance were assessed in a multicenter study that also addressed issues such as observer agreement and the effects of age, arm circumference, heart rate, posture, and blood pressure level on the observer-device differences. We compared 906 simultaneous, same-arm BP measurements in 151 subjects using the TM-2420 versus two skilled clinicians per site using a teaching stethoscope. The agreement between the TM-2420 and mercury column determinations were within 10 mm Hg for 86 to 91% of systolic readings and 91 to 94% of diastolic readings, depending on the posture; a level of agreement which would receive a 'B+' grade from the recent British Hypertension Society guidelines. The limits of agreement (2 standard deviations about the mean difference) for systolic BP between observers and the TM-2420 tended to be greater for the standing position (-20 to 15 mm Hg) compared to supine (-14 to 12 mm Hg) and seated (-13 to 8 mm Hg) positions. Limits of agreement between the observers and device were not dependent upon age, heart rate, arm size, or blood pressure level. Twenty-four-hour blood pressure monitoring in two of the four centers demonstrated an error code rate of 3.4%, excluding 'retries' that are one of the device's features. These data demonstrate an acceptable level of accuracy and performance of the sixth generation of the TM-2420 for use in clinical practice and research. Using multiple centers allowed for a large, heterogenous study population which enabled the assessment of a variety of parameters such as age, heart rate, and posture on the agreement of the device with clinical BP measurements.
更多
查看译文
关键词
AMBULATORY BLOOD PRESSURE MONITORS (ABPM),VALIDATION,CLINICAL EVALUATION,BLOOD PRESSURE MEASUREMENT
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要