Non-Contact Video-Based Neonatal Respiratory Monitoring

Scott L Rossol,Jeffrey K Yang,Caroline Toney-Noland,Janine Bergin, Chandan Basavaraju,Pavan Kumar, Henry C Lee

CHILDREN-BASEL(2020)

引用 18|浏览3
暂无评分
摘要
Respiratory rate (RR) has been shown to be a reliable predictor of cardio-pulmonary deterioration, but standard RR monitoring methods in the neonatal intensive care units (NICU) with contact leads have been related to iatrogenic complications. Video-based monitoring is a potential non-contact system that could improve patient care. This iterative design study developed a novel algorithm that produced RR from footage analyzed from stable NICU patients in open cribs with corrected gestational ages ranging from 33 to 40 weeks. The final algorithm used a proprietary technique of micromotion and stationarity detection (MSD) to model background noise to be able to amplify and record respiratory motions. We found significant correlation-r equals 0.948 (p value of 0.001)-between MSD and the current hospital standard, electrocardiogram impedance pneumography. Our video-based system showed a bias of negative 1.3 breaths and root mean square error of 6.36 breaths per minute compared to standard continuous monitoring. Further work is needed to evaluate the ability of video-based monitors to observe clinical changes in a larger population of patients over extended periods of time.
更多
查看译文
关键词
neonatal monitoring, respiratory rate, clinical alarms, video recording, biomedical technology