Wireless Power Harvesting During Mri

42ND ANNUAL INTERNATIONAL CONFERENCES OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY: ENABLING INNOVATIVE TECHNOLOGIES FOR GLOBAL HEALTHCARE EMBC'20(2020)

Cited 3|Views60
No score
Abstract
We present methods to harvest wireless power directly from the MRI RF field. The system includes a harvester coil to capture RF energy and an RF-DC converter for rectification. Energy harvesting by the harvester coil is modeled as a function of the MRI B-1 RF field. Rectification is modeled using power-dependent large signal S-parameter simulation. A novel reference impedance-based modeling approach is leveraged to cascade models for linear inductive coupling and nonlinear diode rectification, and validated. The method permits independent optimization of harvester coils and RF-DC converters to maximize harvesting efficiency. Feasibility of this technique is demonstrated by implementing concurrent in-bore wireless power harvesting and MRI scanning on a clinical system. The effect of artifacts on image quality is also investigated.
More
Translated text
Key words
Artifacts,Electric Impedance,Magnetic Resonance Imaging