Discrepancy of Trial Rasp and Femoral Stem Relative Position Within the Femoral Canal of a Coated Tapered System: An Intraoperative, Intrapatient Controlled Study.

Arthroplasty today(2020)

引用 1|浏览1
暂无评分
摘要
The quantified stem protrusion appears to be clinically manageable, as only 2 cases required attenuation of stem positioning: in one case by the use of a femoral head with a shorter neck and in the other case by rerasping the femoral bed. Neither case was associated with the most extreme differences in position of the stem with respect to the final rasp. In addition, the used stem shows good overall outcomes in other studies. It appears that factors other than stem and rasp position play a critical role to the surgeon and for total hip arthroplasty success.
更多
查看译文
关键词
Fluoroscopy,Implant position,Radiographic image,Total hip arthroplasty