Comparison of BOLD and CBV using 3D EPI and 3D GRASE for cortical layer fMRI at 7T

Magnetic Resonance in Medicine(2019)

引用 2|浏览11
暂无评分
摘要
Purpose Functional MRI (fMRI) at the mesoscale of cortical layers and columns requires both sensitivity and specificity, which can be compromised if the imaging method is affected by vascular artifacts, particularly cortical draining veins at the pial surface. Recent studies have shown that cerebral blood volume (CBV) imaging is more specific to the actual laminar locus of neural activity than BOLD imaging when using standard gradient-echo (GE) EPI sequences. Gradient and Spin Echo (GRASE) BOLD imaging has also shown greater specificity when compared with GE-BOLD. Methods Here we directly compare CBV and BOLD contrasts in high-resolution imaging of the primary motor cortex for laminar fMRI in four combinations of signal labeling, VASO (CBV) and BOLD with 3D GE-EPI and zoomed 3D GRASE image readouts. Results We find that both CBV imaging using EPI-VASO and BOLD imaging using GRASE-BOLD, show similar specificity and sensitivity and are thus useful tools for mesoscopic fMRI in the human cortex. Conclusion These techniques demonstrate sufficient sensitivity and specificity to allow layer-fMRI to be used by neuroscientists in a wide range of investigations of depth-dependent neural circuitry in the human brain.