Comparison of eight commercial, high-throughput, automated or ELISA assays detecting SARS-CoV-2 IgG or total antibody.

Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology(2020)

引用 57|浏览17
暂无评分
摘要
BACKGROUND:Many commercial assays, of different designs, detecting SARS-CoV-2-specific antibodies exist but with little experience with them. OBJECTIVES:The aim of this study was to compare the performance of assays detecting IgG or total antibodies to N or S antigens, validated for routine use in France, with samples from subjects with more or less severe SARS-CoV-2 infection. METHODS:Eight assays were used: Abbott Architect, DiaSorin Liaison®, bioMérieux Vidas®, Roche Elecsys Cobas®, Siemens Atellica®, BioRad Platelia ELISA, Epitope Diagnostics ELISA, and Wantai ELISA. The tested population included 86 samples from 40 hospitalized subjects and 28 outpatients at different time from symptom onset. RESULTS:The positivity rate varied depending on the assay but was greater for all assays in hospitalized than non-hospitalized patients. Despite a good correlation between the assays, discrepancies occurred, without a systematic origin, even for samples taken more than 20 days after symptom onset. These discrepancies were linked to low antibody levels in pauci-symptomatic patients. CONCLUSION:Whichever assay is chosen, a false negative result may need to be ruled out with another test in a risk situation.