Beam monitor calibration of a synchrotron-based scanned light-ion beam delivery system.

Zeitschrift fur medizinische Physik(2020)

Cited 11|Views50
No score
Abstract
PURPOSE:This paper presents the implementation and comparison of two independent methods of beam monitor calibration in terms of number of particles for scanned proton and carbon ion beams. METHODS:In the first method, called the single-layer method, dose-area-product to water (DAPw) is derived from the absorbed dose to water determined using a Roos-type plane-parallel ionization chamber in single-energy scanned beams. This is considered the reference method for the beam monitor calibration in the clinically relevant proton and carbon energy ranges. In the second method, called the single-spot method, DAPw of a single central spot is determined using a Bragg-peak (BP) type large-area plane-parallel ionization chamber. Emphasis is given to the detailed characterization of the ionization chambers used for the beam monitor calibration. For both methods a detailed uncertainty budget on the DAPw determination is provided as well as on the derivation of the number of particles. RESULTS:Both calibration methods agreed on average within 1.1% for protons and within 2.6% for carbon ions. The uncertainty on DAPw using single-layer beams is 2.1% for protons and 3.1% for carbon ions with major contributions from the available values of kQ and the average spot spacing in both lateral directions. The uncertainty using the single-spot method is 2.2% for protons and 3.2% for carbon ions with major contributions from the available values of kQ and the non-uniformity of the BP chamber response, which can lead to a correction of up-to 3.2%. For the number of particles, an additional dominant uncertainty component for the mean stopping power per incident proton (or the CEMA) needs to be added. CONCLUSION:The agreement between both methods enhances confidence in the beam monitor calibration and the estimated uncertainty. The single-layer method can be used as a reference and the single-spot method is an alternative that, when more accumulated knowledge and data on the method becomes available, can be used as a redundant dose monitor calibration method. This work, together with the overview of information from the literature provided here, is a first step towards comprehensive information on the single-spot method.
More
Translated text
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined