User Experience For Manual Injection Of 2 Ml Viscous Solutions Is Enhanced By A New Prefillable Syringe With A Stacked 8 Mm Ultra-Thin Wall Needle

Aurélie Pager,Anne Combedazou,Karen Guerrero, Tzvetelina Tzvetkova-Chevolleau,Didier Morel,Cécile Frolet,Stanislav Glezer

EXPERT OPINION ON DRUG DELIVERY(2020)

Cited 4|Views3
No score
Abstract
Objectives User experience was compared between a new pre-fillable 2.25 mL glass syringe equipped with an ultra-thin-wall (UTW) 8 mm stacked needle and a marketed BD Neopak (TM) syringe equipped with a special-thin-wall (STW) 12.7 mm stacked needle. Methods Participants simulated subcutaneous injections with both syringes alone (formative Human Factors study) and in combination with a needlestick-prevention device (validation Human Factors study). Results Usability results of both studies showed higher success rates for delivering the full dose of 2 mL viscous solution (30 cP) with the 8mmUTW syringe than with the 12.7mmSTW one (63% vs. 42% in the formative study). The use of the 8mmUTW syringe demonstrated also better ease of use and acceptance results and 72% of formative study participants preferred this new syringe over the current one when delivering the viscous solution. Using a shorter needle also showed a benefit in decreasing the injection-related anxiety. Besides, in the case of a non-recommended injection technique, the calculated risk of accidental intramuscular injection is reduced by 2 to 13 times with the 8mmUTW syringe. Conclusion Altogether, the results obtained demonstrated an improvement of the user experience with this new syringe compared to the current one in the manual delivery of 2 mL viscous solutions.
More
Translated text
Key words
Human factors, usability, user experience, viscous drug delivery, manual subcutaneous injection, prefillable syringe, 8 mm needle, ultra-thin wall cannula, needlestick prevention device, chronic care
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined