Impact of haptic feedback on applied intracorporeal forces using a novel surgical robotic system—a randomized cross-over study with novices in an experimental setup

SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES(2020)

引用 7|浏览8
暂无评分
摘要
Background Most currently used surgical robots have no force feedback; the next generation displays forces visually. A novel single-port robotic surgical system called FLEXMIN has been developed. Through an outer diameter of 38 mm, two instruments are teleoperated from a surgeon’s control console including true haptic force feedback. One additional channel incorporates a telescope, another is free for special instrument functions. Methods This randomized cross-over study analyzed the effect of haptic feedback on the application of intracorporeal forces. In a standardized experiment setup, the subjects had to draw circles with the surgical robot as gently as possible. The applied forces, the required time spans, and predefined error rates were measured. Results Without haptic feedback, the maximum forces (median/IQR) were 6.43 N/2.96 N. With haptic feedback, the maximum forces were lower (3.57 N/1.94 N, p < 0.001). Also, the arithmetic means of the force progression ( p < 0.001) and their standard deviations ( p < 0.001) were lower. Not significant were the shorter durations and lower error rates. No sequence effect of force or duration was detected. No characteristic learning or fatigue curve was observed. Conclusions In the experiment setup, the true haptic force feedback can reduce the applied intracorporeal robotic force to one-half when considering the aspects maximum, means, and standard deviation. Other test tasks are needed to validate the influence of force feedback on surgical efficiency and safety.
更多
查看译文
关键词
Surgical robot, Haptics, Force feedback, Randomized controlled trial, Experiment setup
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要