GATE Simulation of a Clinical PET Scanner : Influence of Windows Timing Coincidences and Dead Time on Count Rate Performance

semanticscholar(2017)

引用 0|浏览4
暂无评分
摘要
This paper present an evaluation of the Count rate performance and noise equivalent count rate (NECR) in different coincidence timing windows arranging from 8 to 12 ns, with different dead-time electronic 5000ns paralyzable and 4900ns nonparalyzable. The clinical PET ECAT EXACT HR+ is validated according to the National Electrical Manufacturers Association (NEMA NU 2-2012). The simulated results (scatter fraction (SF), sensitivity(S) and NECR) show a good agreement with the experimental data. the results show that decreasing the coincidence time windows cause a decrease of 28% of the NECR. Moreover, the NECR is increased by 4,28% using 4900ns non paralyzable deadtime instead of 5000ns parlyzable. Keywords—PET, scatter fraction, sensitivity, NECR, dead time