On contrast two-dimensional echocardiography : applications and new developments . part viii *

semanticscholar(2010)

Cited 0|Views0
No score
Abstract
To investigate whether precision microbubbles can be simply generated and used for quantitative contrast echocardiographic studies, precision microbubbles were fabricated in gelatin using a simple generator. The generator consisted of a fluid pump, a generating site containing a narrow bore tube and a bubble stream output port. Up to 5,000 microbubbles/s were generated with a controllable radius of 80 to 150 11-. No detectable interbubble size variation was observed under high power microscopy. To examine whether these bubbles could be used for quantitative contrast echocardiographic studies, they were then infused into a tube with a known flow of degassed water during ultrasonic imaging. The resulting ultrasonic contrast effect in the tube lumen at
More
Translated text
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined