We evaluated the performance of the BioFire® Respiratory Panel 2.1 (RP2.1) in the detection of SARS CoV-2 in comparison against three other SARS CoV-2 EUA assays. In these studies, the RP2.1 panel had 98 % positive percent agreement (48/49) and 100 % negative percent agreement (49/49). Since 30 % of nasopharyngeal swab specimens have a SARS CoV-2 Ct >30 and thus detection of virus in low titers is clinically relevant, a sample with a high titer was diluted and each 10 fold dilution was tested in triplicate and compared against 6 other EUA approved SARS CoV-2 assays. These data suggested that the BioFire® RP2.1 panel, along with four other SARS CoV-2 assays (Roche cobas, Cepheid Xpert Xpress, BioFire® Defense COVID19, and NECoV19), consistently detected viral RNA at the 10-7 dilution. Overall, these studies suggest that the BioFire® RP2.1 assay can be used to detect acute cases of SARS CoV2 in addition to patients with low viral titer later in disease presentation.