谷歌浏览器插件
订阅小程序
在清言上使用

Monitoring patient movement with boron neutron capture therapy and motion capture technology.

Applied Radiation and Isotopes(2020)

引用 3|浏览18
暂无评分
摘要
In boron neutron capture therapy (BNCT), a patient must remain in a fixed position during the irradiation process. In this study, a system was devised that can guide a patient to the correct position and the patient can be monitored during the irradiation process. This is achieved by using motion capture technology that consists of many cameras. The discrepancy of the measured coordinates for each marker on a phantom by the system was less than 5 mm. For practical applications, further research and verification are required.
更多
查看译文
关键词
Patient positioning,BNCT,Motion capture technology,Real-time monitoring,Patient immobilization
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要