The transcutaneous electromyography recording method for intraoperative neuromonitoring of recurrent laryngeal nerve during minimally invasive parathyroidectomy

SCIENTIFIC REPORTS(2020)

Cited 6|Views7
No score
Abstract
Intraoperative neuromonitoring (IONM) facilitates recurrent laryngeal nerve (RLN) protection in thyroid and parathyroid surgeries. This study aimed to investigate a novel transcutaneous electromyography (EMG) recording method for IONM of the RLN during minimally invasive parathyroidectomy (MIP). Twenty patients with primary hyperparathyroidism undergoing MIP were enrolled. Two paired needle electrodes were percutaneously inserted into the bilateral laminas of thyroid cartilage for monitoring the vagus nerve and RLN. A standardized IONM procedure (V 1 -R 1 -R 2 -V 2 signals) was strictly followed, and the RLN was routinely located and mapped. Pre- and postoperative laryngofiberoscopy was performed to confirm vocal cord function. The proposed technique was successfully used in all patients, and typical EMG signals were effectively detected. No significant change in EMG signals before and after tumor resection was noted, and a normal vocal cord movement was ensured in all patients with postoperative laryngofiberoscopy. IONM helped localize the position of the RLN and facilitated the safe resection of the parathyroid tumor during MIP. The novel transcutaneous EMG recording method proposed in this study was feasible, convenient, reliable, and inexpensive.
More
Translated text
Key words
Action potential generation,Parathyroid diseases,Science,Humanities and Social Sciences,multidisciplinary
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined