Effectiveness Of Robust Optimization In Volumetric Modulated Arc Therapy Using 6 And 10 Mv Flattening Filter-Free Beam Therapy Planning For Lung Stereotactic Body Radiation Therapy With A Breath-Hold Technique

JOURNAL OF RADIATION RESEARCH(2020)

引用 3|浏览6
暂无评分
摘要
We investigated the feasibility of a robust optimization with 6 MV X-ray (6X) and 10 MVX-ray (10X) flattening filter-free (FFF) beams in a volumetric modulated arc therapy (VMAT) plan for lung stereotactic body radiation therapy (SBRT) using a breath-holding technique. Ten lung cancer patients were selected. Four VMAT plans were generated for each patient; namely, an optimized plan based on the planning target volume (PTV) margin and a second plan based on a robust optimization of the internal target volume (ITV) with setup uncertainties, each for the 6X- and 10X-FFF beams. Both optimized plans were normalized by the percentage of the prescription dose covering 95% of the target volume (D-95%) to the PTV (1050 cGy x 4 fractions). All optimized plans were evaluated using perturbed doses by specifying user-defined shifted values from the isocentre. The average perturbed D-99(%) doses to the ITV, compared to the nominal plan, decreased by 369.1 (6X-FFF) and 301.0 cGy (10X-FFF) for the PTV-based optimized plan, and 346.0 (6X-FFF) and 271.6 cGy (10X-FFF) for the robust optimized plan, respectively. The standard deviation of the D-99(%) dose to the ITV were 163.6 (6X-FFF) and 158.9 cGy (10X-FFF) for the PTV-based plan, and 138.9 (6X-FFF) and 128.5 cGy (10X-FFF) for the robust optimized plan, respectively. Robust optimized plans with 10X-FFF beams is a feasible method to achieve dose certainty for the ITV for lung SBRT using a breath-holding technique.
更多
查看译文
关键词
robust optimization, lung, VMAT, dose uncertainty, perturbed dose
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要