Accuracy and stability of an arterial sensor for glucose monitoring in a porcine model using glucose clamp technique

SCIENTIFIC REPORTS(2020)

引用 2|浏览16
暂无评分
摘要
Intravascular glucose sensors have the potential to improve and facilitate glycemic control in critically ill patients and might overcome measurement delay and accuracy issues. This study investigated the accuracy and stability of a biosensor for arterial glucose monitoring tested in a hypo- and hyperglycemic clamp experiment in pigs. 12 sensors were tested over 5 consecutive days in 6 different pigs. Samples of sensor and reference measurement pairs were obtained every 15 minutes. 1337 pairs of glucose values (range 37–458 mg/dl) were available for analysis. The systems met ISO 15197:2013 criteria in 99.2% in total, 100% for glucose <100 mg/dl (n = 414) and 98.8% for glucose ≥100 mg/dl (n = 923). The mean absolute relative difference (MARD) during the entire glycemic range of all sensors was 4.3%. The MARDs within the hypoglycemic (<70 mg/dl), euglycemic (≥70–180 mg/dl) and hyperglycemic glucose ranges (≥180 mg/dl) were 6.1%, 3.6% and 4.7%, respectively. Sensors indicated comparable performance on all days investigated (day 1, 3 and 5). None of the systems showed premature failures. In a porcine model, the performance of the biosensor revealed a promising performance. The transfer of these results into a human setting is the logical next step.