A Bridged Loop Gap Resonator (Blgr) For Small Animal Imaging By 1.5 T Mri Systems

C Staat, M Mützel,A Haase

REVIEW OF SCIENTIFIC INSTRUMENTS(2020)

引用 0|浏览2
暂无评分
摘要
A bridged loop gap resonator (BLGR) was developed as a transmit and receive coil for a mobile insert to be used for small animal proton imaging by 1.5 T MRI devices. The insert system has its own gradient system, radio frequency (RF) transmit and receive coil, and control and signal processing unit. The reflection S-11 and transmission S-21 parameters, quality factor (Q), sensitivity, signal to noise ratio (SNR), and maps of the static (B-0) and RF (B-1) magnetic flux densities were measured. The RF coil was developed starting from a loop gap resonator (LGR) for a balanced LGR and a shielded balanced LGR for a shielded bridged balanced LGR. The purpose of developing this device is to minimize the influence of the sample and surroundings on the RF coil parameters. The final design of the BLGR does not require retuning after a sample change. A 3D image of a mouse in formalin was acquired with a fast low angle shot (FLASH) MRI sequence. The SNR was calculated from one FLASH image. The signal for SNR calculation was acquired from a gadolinium-doped water sample and the noise from the air outside of the sample. This article verifies that the BLGR is viable for small animal nuclear magnetic resonance imaging at 1.5 T and is independent of sample size and material. Published under license by AIP Publishing.