Recent Least Burdensome Approach For The Approval Of Innovative Medical Devices In Japan -Regulatory Approval Review Of An Everolimus-Eluting Bioresorbable Scaffold-

INTERNAL MEDICINE(2021)

引用 0|浏览13
暂无评分
摘要
Although a domestic trial in Japan revealed that Absorb bioresorbable vascular scaffold (BVS) has no inferiority to everolimus-eluting stent (EES) cohort in the primary endpoint of the target lesion failure at 12 months, the scaffold/stent thrombosis (ST) rates with the BVS at 24 months were higher than those with the EES (Absorb BVS 3.1% vs. EES 1.5%), the ST rate of 3.1% with Absorb BVS is not an acceptable level in Japan. A cause-of-ST analysis revealed that cases in which diagnostic imaging and ensuing post-dilatation had been performed appropriately had lower ST rates than those without such management (within 1 year: 1.37% vs. 7.69%, from 1 to 2 years: 0.00% vs. 8.33%). Therefore, a further evaluation was needed to confirm that the ST rate with the Absorb BVS would be reduced by a proper implementation procedure. Regulatory approval was given conditionally to initiate rigorous post-marketing data collection in order to ensure the proper use of this device in limited facilities. The One-year Use-Result Survey in Japan for the Absorb BVS revealed no instances of ST. This approach to reducing the premarket regulatory burden of clinical trials and enhancing the post-marketing commitments of medical device regulation is useful for expediting patient access to innovative medical devices.
更多
查看译文
关键词
everolimus-eluting bioresorbable scaffold, Use-Result Survey, scaffold thrombosis