A low-cost smartphone controlled portable system with accurately confined on-chip 3D electrodes for flow-through cell electroporation.

Bioelectrochemistry (Amsterdam, Netherlands)(2020)

引用 6|浏览33
暂无评分
摘要
Microscale flow-through electroporation at DC voltage has advantages in delivering small molecules. Yet, electroporation based on constant voltage are liable to generate electrolysis products which limits the voltage-operating window. Parallel on-chip 3D electrodes with close and uniform spacing are required to cut down voltage as well as provide enough electric field for electroporation. Here we present a simple electrode fabrication method based on capillary restriction valves in Z-axis to achieve parallel 3D electrodes with controllable electrode spacing in PDMS chips. With electrodes accurately placed in close range, a low voltage of only 1.5 V can generate enough electric field (>400 V/cm) to make cell membrane permeable. Squeeze flow is introduced to produce higher electric field (>800 V/cm) at a fixed voltage for more efficient electroporation. Benefit from the electrode fabrication method and application of squeeze flow, we develop a smartphone controlled microfluidic electroporation system which integrate functions of sample injection, pressure regulating, real-time observation and constant DC power supply. The system is used to electroporate two cell lines, showing a permeabilization percentage of 63% for HEK-293 cells and 58% for CHO-K1 cells with optimal parameters. Thus, the portable microfluidic system provides a cost-effective and user-friendly flow-through cell electroporation platform.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要