GAMMA SPECTRUM STABILIZATION FOR ENVIRONMENTAL RADIATION MONITORING STATIONS USING NAI(TL) DETECTOR.

RADIATION PROTECTION DOSIMETRY(2020)

Cited 5|Views6
No score
Abstract
Gamma spectrum measured by an NaI(Tl) detector is known to be unstable with the in situ temperature. In the present work, an advanced method has been applied to stabilize the gamma spectrum measured by the NaI(Tl) detector at environmental radiation monitoring (ERM) stations. The method is based on experimental data obtained under controlled conditions in laboratory. In the temperature range from 4 to 45 degrees C, the relative deviation of the peak positions within the stabilized gamma spectrum is less than 2%. To test this method in a real scenario, it has been integrated into the ERM station at the Military Institute of Chemical and Environmental Engineering in Hanoi, Vietnam. The results show that the proposed method is ready for a real application.