Initial Evaluation of the Performance of Novel Inorganic Scintillating Detectors for Small Animal Irradiation Dosimetry

IEEE Sensors Journal(2020)

引用 14|浏览9
暂无评分
摘要
The purpose of this study was to design and evaluate the performance of four novel inorganic scintillating detectors (ISDs) on the Small Animal Radiation Research Platform (SARRP). Relative scintillator output, measurement repeatability, setup uncertainty, linearity with dose rate, and signal reproducibility over time were investigated. The Gd 2 O 2 S:Tb detector had the highest relative signal output, generating up to 219 times more charge than a previously characterized BCF-60-based plastic scintillating detector (PSD). The Gd 2 O 2 S:Tb detector was then used to measure 220 kVp therapy beam profiles of 10 $\times$ 10 and 5 $\times$ 5 mm 2 fields. Beam profiles using the ZnS-based phosphor were also obtained and compared to investigate the performance of a lower density inorganic scintillator. 10 $\times$ 10 and 5 $\times$ 5 mm 2 therapy beam profile measurements made with the Gd 2 O 2 S:Tb and BCF-60 detectors differed, on average, by 1.1% and 1.9%, respectively. The ZnS:Ag measurements differed, on average, by 2.5% and 6% relative to BCF-60 measurements of the 10 $\times$ 10 and 5 $\times$ 5 mm 2 beam profiles, respectively. MicroCT imaging of the detector volumes was also performed, revealing poor packing of the ZnS:Ag crystalline phosphor in the deepest region of the cylindrical cavity. The Gd 2 O 2 S:Tb detector, in particular, has proven to be a promising candidate for real-time dosimetry of small fields in small animal irradiators, primarily because of the very large signal intensities observed, along with good repeatability, dose rate linearity, reproducibility and agreement with beam profile measurements made with a previously validated detector.
更多
查看译文
关键词
Dosimetry,optical fibers,radiation detectors,scintillators
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要