Quantitative amino acid analysis by liquid chromatography-tandem mass spectrometry using low cost derivatization and an automated liquid handler.

JIMD reports(2020)

Cited 12|Views6
No score
Abstract
Amino acid analysis is central to newborn screening and the investigation of inborn errors of metabolism. Ion-exchange chromatography with ninhydrin derivatization remains the reference method for quantitative amino acid analysis but offers slow chromatography and is susceptible to interference from other co-eluting compounds. Liquid-chromatography tandem mass spectrometry (LC-MS/MS) provides a rapid and highly specific alternative, but sample preparation is frequently laborious and sometimes cost prohibitive. To address these limitations, we validated an LC-MS/MS method using the aTRAQ Reagents Application Kit with a modified protocol consuming only half reagents. Adequate performance for clinical specimen measurement of 26 amino acids with high clinical relevance was achieved. An automated liquid handler and modified calibration and normalization approaches were used to ensure reproducible assay performance. Linear measurement between 5 and 2000 μM was achieved for most analytes despite use of a small, 10 μl sample size. Overall the method achieved near substantially improved throughput and enabled use of smaller samples volumes for batched analyses of clinical samples.
More
Translated text
Key words
amino acid,inborn errors of metabolism,isotope‐coded derivatization,liquid chromatography,mass spectrometry,triple quadrupole
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined