Circuit Design Challenges of ADC for the Application in Multiple Physiological Parameters Detection System

2019 IEEE 13th International Conference on ASIC (ASICON)(2019)

引用 2|浏览4
暂无评分
摘要
Design challenges of ADC for the application in multiple physiological parameters detection system is discussed in this paper. Based on requirements in the physiological signal detection applications, design difficulties of these high precision Sigma-Delta ADCs, including multi-channel high effective resolution, ultra-low power and low channel mismatch, are analyzed with design tradeoff considerations. Results of a Sigma-Delta modulator for physiological parameters detection applications is given. The noise floor of the modulator is about -120dB, the SNR is 101.4dB, and ENOB is 16.55 bit, with the power consumption of 18 mW.
更多
查看译文
关键词
multiple physiological parameters detection system,physiological signal detection applications,design difficulties,multichannel high effective resolution,low channel mismatch,physiological parameters detection applications,high precision Sigma-Delta ADC,noise figure 120.0 dB,noise figure 101.4 dB,word length 16.55 bit,power 18.0 mW