Development of a realistic 3D printed eye lens dosemeter using CAD integrated with Monte Carlo simulation

BIOMEDICAL PHYSICS & ENGINEERING EXPRESS(2020)

引用 4|浏览10
暂无评分
摘要
Recent epidemiological studies suggested to lower the threshold dose for radiation induced cataract in the eye lens. Therefore, eye lens radiation protection became to play a more important role in personal dosimetry. The main objective of this work is to propose a new methodology for prototyping and benchmarking of an eye lens dosimter based on the equivalent dose to the sensitive part of the eye lens, using CAD Software and Geant4 Monte Carlo simulations with mesh modelling and 3D printing. A 3D printed dosemeter was type tested based on IEC 62387:2012, in terms of energy and angular dependence for the measurements of Hp(3). The results show that the methodology employed is suitable for the development of new eye lens dosemeters.
更多
查看译文
关键词
eye lens radiation monitoring,3D-printing,personnel dosimetry