谷歌浏览器插件
订阅小程序
在清言上使用

Inexpensive Robotic System For Standard And Fluorescent Imaging Of Protein Crystals

Dakota Handzlik,Eric T. Larson, Erika Munschy, Galina Obmolova, Delphine Collin,Timothy K. Craig

ACTA CRYSTALLOGRAPHICA SECTION F-STRUCTURAL BIOLOGY COMMUNICATIONS(2019)

引用 2|浏览6
暂无评分
摘要
Protein-crystallization imaging and classification is a labor-intensive process typically performed either by humans or by instruments that currently cost well over $100 000. This cost puts the use of crystallization-trial imaging outside the reach of most academic laboratories, and also start-up biotechnology firms, where resources are scarce. An imaging system has been designed and prototyped which automatically captures images from multi-well protein-crystallization experiments using both standard and fluorescent imaging techniques at a cost 28 times lower than current market rates. The machine uses a Panowin F1 3D printer as a base and controls it using G-code commands sent from a Python script running on a desktop computer. A graphical user interface (GUI) was developed to enable users to control the machine and facilitate image capture, classification and editing. A 488 nm laser diode and a 525 nm filter were incorporated to allow in situ fluorescent imaging of proteins trace-labeled with a fluorophore, Alexa Fluor 488. The instrument was primarily designed using a 3D printer and augmented using commercially available parts, and this publication aims to serve as a guide for comparable in-laboratory robotics projects.
更多
查看译文
关键词
fluorescent imaging,automated microscope,imaging system,open source,robotics
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper