Low-pressure TPC with THGEM readout for ion identification in Accelerator Mass Spectrometry

A. Bondar, A. Buzulutskov, V. Parkhomchuk, A. Petrozhitskiy, T. Shakirova, A. Sokolov

NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT(2020)

Cited 3|Views15
No score
Abstract
A new technique for ion identification in Accelerator Mass Spectrometry (AMS) has been proposed based on measuring the ion track ranges using a low-pressure TPC. As a proof of principle, a low-pressure TPC with charge readout using a THGEM multiplier was developed. The tracks of alpha particles from various radioactive sources were successfully recorded in the TPC. The track ranges were measured with a rather high accuracy, reaching 2%. Using these results and SRIM code simulation, it is shown that the isobaric boron and beryllium ions can be effectively separated at the 10 sigma level. This technique is expected to be applied in the AMS facility in Novosibirsk for dating geological objects, in particular for geochronology of Cenozoic Era.
More
Translated text
Key words
Low-pressure TPC,THGEM readout,Accelerator mass spectrometry,Ion identification