[Value of cone-beam computer tomography in the middle and inner ear observation].

Lin chuang er bi yan hou tou jing wai ke za zhi = Journal of clinical otorhinolaryngology, head, and neck surgery(2021)

Cited 0|Views5
No score
Abstract
To identified the feasibility and normal range of cone beam computer tomography(CBCT) in the measurement of temporal bone. 15 formalin fixed human cadaver head specimens were scanned by CBCT, high resolution CT, and Micro CT, respectively. Morphological parameter measurements of the middle and inner ear structures including ossicular chain, cochlea, semicircular canal and facial nerve were performed, and the results measured by the three scanning methods were compared. None of the parameters measured by the three scanning methods were statistically significant except the thickness of stapes footplate(<0.01) and the diameter of cochlear basal turn(<0.01). CBCT was superior in detecting facial nerve bony canal dehiscence. CBCT has the advantages of short scanning time, low radiation dose and high resolution. It can accurately display the morphological characteristics of the temporal bone structures, and is a reliable evaluation method for otological surgery.
More
Translated text
Key words
cone-beam computer tomography,multiple plane reconstruction,temporal bone measurement