Validation of an Affordable Handheld Wavefront Autorefractor.

OPTOMETRY AND VISION SCIENCE(2019)

引用 19|浏览26
暂无评分
摘要
PURPOSE The purpose of this study was to validate the commercial version of a portable wavefront autorefractor for measuring refractive errors. METHODS Refraction was performed without cycloplegia using (1) a standard clinical procedure consisting of an objective measurement with a desktop autorefractor followed by subjective refraction (SR) and (2) with the handheld autorefractor. Agreement between both methods was evaluated using Bland-Altman analysis and by comparing the visual acuity (VA) with trial frames set to the resulting measurements. RESULTS The study was conducted on 54 patients (33.9 +/- 14.1 years of age) with a spherical equivalent (M) refraction determined by SR ranging from -7.25 to 4.25 D (mean +/- SD, -0.93 +/- 1.95 D). Mean differences between the portable autorefractor and SR were 0.09 +/- 0.39, -0.06 +/- 0.13, and 0.02 +/- 0.12 D for M, J(0), and J(45), respectively. The device agreed within 0.5 D of SR in 87% of the eyes for spherical equivalent power. The average VAs achieved from trial lenses set to the wavefront autorefractor and SR results were 0.02 +/- 0.015 and 0.015 +/- 0.042 logMAR units, respectively. Visual acuity resulting from correction based on the device was the same as or better than that achieved by SR in 87% of the eyes. CONCLUSIONS This study found excellent agreement between the measurements obtained with the portable autorefractor and the prescriptions based on SR and only small differences between the VA achieved by either method.