谷歌浏览器插件
订阅小程序
在清言上使用

Proximal outflow graft extrinsic compression of the Heartmate 3 centrifugal flow left ventricular assist device resulting in pump dysfunction

Journal of Artificial Organs(2020)

引用 3|浏览15
暂无评分
摘要
We herein report an unusual life-threatening complication with the Hearmate 3 centrifugal ventricular assist device: proximal outflow graft extrinsic compression due to external thrombus leading to pump dysfunction and urgent heart transplantation. Low flow alarms without other abnormalities lead to the diagnosis of outflow occlusion. There was no evidence for a twist of the outflow graft, and no evidence for outflow thrombosis. The location of the outflow occlusion was particularly difficult to diagnose due to metallic artifacts around the pump. This report underscores that this complication should be screened in case of persistent low-flow alarms with the Heartmate 3 centrifugal pump.
更多
查看译文
关键词
Left ventricular assist device,Complication,Thrombosis
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要