Construction method through multiple off-axis parabolic surfaces expansion and mixing to design an easy-aligned freeform spectrometer.

OPTICS EXPRESS(2019)

引用 17|浏览15
暂无评分
摘要
The classic Czerny-Turner spectrometer consists of a plane grating and two spherical mirrors. The optical path geometry adopted for incident and grating dispersed light is off-axis reflection, so the spherical collimating and focusing mirrors introduce coma and astigmatism. The conventional configuration is asymmetrical for coma automatic compensation, but suffers from astigmatism. We substitute the off-axis parabolic (OAP) surfaces for spherical surfaces of the collimating mirror and each sub-region of the focusing mirror, to achieve an aberration free configuration. The multiple OAP surfaces are then expanded and mixed, to construct a freeform surface integrating the collimating and focusing mirrors into a single clement. Results show that a 0.1 nm spectral resolution is achieved over a bandwidth of 400 nm centered at 800 nm, in the designed spectrometer comprised of a plane grating and one freeform mirror. The construction method is advantageous to integrated optic design, and the resulting freeform mirror spectrometer is compact, and simplifies manufacture and alignment. (C) 2019 Optical Society of America under the terms of the OSA Open Access Publishing Agreement
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper