Interval analysis of blood filtration efficacy in novel device for autotransfusion in postpartum hemorrhage.

Journal of medical engineering & technology(2019)

Cited 0|Views16
No score
Abstract
Postpartum haemorrhage (PPH), the leading cause of maternal mortality, is particularly problematic in low resource settings where access to safe blood supplies and definitive medical treatment is limited. We describe the continued development of an autotransfusion device designed to treat PPH by collection, filtration and infusion of maternal blood. Previous study has demonstrated that the device effectively moves blood through a filtration apparatus and removes up to 97% of aerobic bacteria but had poor anaerobic bacteria reduction. In this study, we investigate the filtration efficacy of the device using configurations comprised of three different leukocyte depletion filter designs: the Pall Leukoguard RS leukocyte reduction filter (PLRF), the Haemonetics BPF4™ (BPF4) leukocyte reduction filter, and the Haemonetics SCRC Leukotrap® (SCRC) filter. All configurations performed well with reductions ranging from 49 to 98%. Configurations containing 2 Haemonetics SCRC Leukotrapfilters (configuration 5 and 6) consistently reduced anaerobic bacteria by at least 73%. These results indicate that utilising a combination of SCRC and PLRF filters confers a high level of microbial filtration with improved removal of anaerobic organisms.
More
Translated text
Key words
Postpartum haemorrhage,global health,medical device,microbial filtration,obstetrics