Handheld projective imaging device for near-infrared fluorescence imaging and intraoperative guidance of sentinel lymph node resection.

JOURNAL OF BIOMEDICAL OPTICS(2019)

引用 9|浏览11
暂无评分
摘要
We propose a handheld projective imaging device for orthotopic projection of near-infrared fluorescence images onto target biological tissue at visible wavelengths without any additional visual aid. The device integrates a laser diode light source module, a camera module, a projector, an ultrasonic distance sensor, a Raspberry Pi single-board computer, and a battery module in a rugged handheld unit. It is calibrated at the detected working distance for seamless coregistration between fluorescence emission and projective imaging at the target tissue site. The proposed device is able to achieve a projection resolution higher than 314 mu m and a planar projection bias less than 1 mm at a projection field of view of 58 x 108 mm(2) and a working distance of 27 cm. Technical feasibility for projective imaging is verified in an ex vivo model of chicken breast tissue using indocyanine green as a fluorescence agent. Clinical utility for image-guided surgery is demonstrated in a clinical trial where sentinel lymph nodes in breast cancer patients are identified and resected under the guidance of projective imaging. Our ex vivo and in vivo experiments imply the clinical utility of deploying the proposed device for image-guided surgical interventions in resource-limited settings. (C) The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License.
更多
查看译文
关键词
handheld,near-infrared fluorescence imaging,projective imaging,surgical navigation,sentinel lymph node