Oversensing issues leading to device extraction: When subcutaneous implantable cardioverter-defibrillator reached a dead-end

Heart rhythm(2020)

Cited 19|Views39
No score
Abstract
BACKGROUND Subcutaneous implantable cardioverter-defibrillator (S-ICD) implantations are rapidly expanding. However, the subcutaneous detection and interpretation of cardiac signals in S-ICDs is much more challenging than by conventional devices. There is a complete paradigm shift in cardiac signal sensing with subcutaneous signal detection, leading in some cases to oversensing with restricted programming options. OBJECTIVES The aim of this single-center study was to quantify and describe cases where recurring oversensing made the extraction of the device necessary. METHODS Consecutive patients (n = 108) implanted with an S-ICD in our tertiary referral hospital were considered for analysis. Clinical and remote monitoring data were analyzed. RESULTS The S-ICD had to be explanted in 6 of 108 implanted patients (5.6%) because of refractory oversensing issues: myopotential oversensing, P- or T-wave oversensing, rate-dependent left bundle branch block aberrancy during exercise with R-wave double counting, and R-wave amplitude decrease after ventricular tachycardia ablation leading to noise detection. Seventeen of 108 patients experienced oversensing (15.7%): 9 patients had at least 1 inappropriate charge without a shock (8.3%), 3 patients had at least 1 inappropriate shock (2.8%), and 5 patients had both episodes (4.6%). CONCLUSION So far, cardiologists have had to deal with transvenous ICD lead fractures, but signal oversensing without correcting programming option could be the equivalent weakness of S-ICDs, despite an adequate screening.
More
Translated text
Key words
Implantable cardioverter-defibrillator,Inappropriate therapy,Oversensing,Subcutaneous implantable cardioverter-defibrillator,Sudden cardiac death
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined