OBJECTIVE: We report a case of late type IIIb endoleak with Willis covered stent (WCS) developed 14 months after endovascular paraclinoid aneurysm repair. METHODS: A 52-year-old woman presented with episodic headache, caused by a giant paraclinoid aneurysm. She underwent a successful 3.5 x 16mm WCS positioning to treat the aneurysm. Fourteen months later, the patient was admitted with the same symptoms. Digital subtraction angiography examination showed recurrence of the aneurysm, which was similar to the preoperative one. DynaCT (Siemens, Erlangen, Germany) indicated the intact of the metal structure of the stent without migration. Type IIIb endoleak (defect in the graft fabric) was confirmed with a whole aneurysm neck located in the middle part of the stent. The type IIIb endoleak was treated with another WCS (4.0 x 16mm). The immediate digital subtraction angiography imaging indicated that the endoleak disappeared and the aneurysm was completely occluded. Re-examination done 1 year after the second treatment showed a complete exclusion of the aneurysm sac. CONCLUSIONS: Type IIIb endoleaks can be safely treated by the endovascular positioning of another WCS. Continuous surveillance after endovascular paraclinoid aneurysm repair for intracranial aneurysms is warranted to make ensure the safety of WCS.