Design of a PGNAA facility using D-T neutron generator for bulk samples analysis

Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms(2019)

引用 11|浏览22
暂无评分
摘要
A prompt gamma-ray neutron activation analysis (PGNAA) facility has been developed for the determination of elements in bulk samples. The basic design of the PGNAA facility is composed of three major components: A deuterium-tritium neutron generator, the neutron moderator and reflector, and a high-purity germanium detector. The Monte Carlo code was used to select materials of moderator and optimize geometrical size of the facility. Then the PGNAA facility was built based on the simulation results. The bulk aqueous samples containing chlorine and cadmium were analyzed with the facility. The neutron self-shielding effect of the bulk aqueous samples was also corrected. The results showed the minimum detectable concentration (MDC) of chlorine and cadmium were 2.58 ± 0.78 g/L and 0.05 ± 0.01 g/L, respectively. For mixture sample analysis, the relative deviations of calculated value and actual value for chlorine and cadmium were 5.1% and 2.0%, respectively.
更多
查看译文
关键词
PGNAA,D-T neutron generator,Bulk sample analysis,Neutron self-shielding correction