Assessment of the Bacharach Hi Flow® Sampler Characteristics and Potential Failure Modes when Measuring Methane Emissions

J.I. Connolly, R.A. Robinson, T.D. Gardiner

Measurement(2019)

引用 22|浏览1
暂无评分
摘要
•Sampler suitable for measuring methane emissions if operated within its limitations.•Both sensor modes of the Hi Flow Sampler show good linearity (R2 > 0.992).•Noisy mode switch region around 5.9% methane by volume which can be avoided by adjusting flow rate.•Failure due to non-methane hydrocarbons only occurs sometimes.•Recommend to calibrate weekly, verify daily and shut off between measurements.
更多
查看译文
关键词
Methane emissions,Bacharach Hi Flow Sampler,Natural Gas quantification,Instrument performance,Methane leaks