In vitro validation of indirect calorimetry device developed for the ICALIC project against mass spectrometry

Clinical Nutrition ESPEN(2019)

引用 20|浏览16
暂无评分
摘要
In vitro evaluation confirms that the accuracy of the Q-NRG indirect calorimeter is within 5% at oxygen enrichment to 70%; i.e. maximum expected for clinical use. Further recommendations for the clinical use of the Q-NRG by will be released once the ongoing multi-center study is completed.
更多
查看译文
关键词
Indirect calorimeter,VO2,VCO2,Critically ill,Mechanical ventilation,Validation