Corrosion Generation and Cleaning Effect on Surgical Instruments with Attached Radiofrequency Identification Tags in Long-Term Usage.

SURGICAL INFECTIONS(2019)

引用 4|浏览10
暂无评分
摘要
Background: Faulty cleaning of surgical instruments may lead to corrosion damage and a higher risk of surgical site infection. We have developed a method in which each instrument has an attached radiofrequency identification (RFID) tag for individual management. However, because of the structure of the instruments, a risk of corrosion from poor cleaning exists; therefore, observation during long-term usage is necessary. Methods: The cleaning effect at the jig of the RFID tag was verified by the amount of residual protein left by various cleaning methods. In our investigation of long-term usage, we examined 94 surgical instruments with RFID tags used in the operating room for 50 months employing a microscope to identify any corrosion at the jig. Results: The method using a washer disinfector (WD) was found to be highly effective. From observation after long-term usage, friction by the RFID tag occurred in about 70% of the jigs. However, no pitting or general corrosion was seen. Conclusions: When WD is used properly, there is only a minor risk of residual protein, and corrosion does not occur even with long-term use. By using surgical instruments with RFID tags, it is possible to determine the number of uses and the history at the individual level. This facilitates operation of safe surgical instruments by limiting the number of times a particular instrument is used.
更多
查看译文
关键词
corrosion,long-term usage,residual protein,RFID tags,surgical instrument