A dual-tuned multichannel bilateral RF coil for 1 H/ 23 Na breast MRI at 7 T.

MAGNETIC RESONANCE IN MEDICINE(2019)

引用 16|浏览20
暂无评分
摘要
Purpose: Sodium MRI has shown promise for monitoring neoadjuvant chemotherapy response in breast cancer. The purpose of this work was to build a dual-tuned bilateral proton/sodium breast coil for 7T MRI that provides sufficient SNR to enable sodium breast imaging in less than 10 minutes. Methods: The proton/sodium coil consists of 2 shielded unilateral units: 1 for each breast. Each unit consists of 3 nested layers: (1) a 3-loop solenoid for sodium excitation, (2) a 3-loop solenoid for proton excitation and signal reception, and (3) a 4-channel receive array for sodium signal reception. Benchmark measurements were performed in phantoms with and without the sodium receive array insert. In vivo images were acquired on a healthy volunteer. Results: The sodium receive array boosted 1.5 to 3 times the SNR compared with the solenoid. Proton SNR loss due to residual interaction with the sodium array was less than 10%. The coil enabled sodium imaging in vivo with 2.8-mm isotropic nominal resolution (similar to 5-mm real resolution) in 9:36 minutes. Conclusion: The coil design that we propose addresses challenges associated with sodium's low SNR from a hardware perspective and offers the opportunity to investigate noninvasively breast tumor metabolism as a function of sodium concentration in patients undergoing neoadjuvant chemotherapy.
更多
查看译文
关键词
bilateral breast coil,breast MRI,high-field MRI,multi-nuclei coil,sodium (Na-23) MRI