Modelling A Transition-Edge Sensor X-Ray Microcalorimeter Linear Array For Compton Profile Measurements And Energy Dispersive Diffraction

IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY(2019)

引用 2|浏览109
暂无评分
摘要
Transition-edge sensors (TES) are a type of superconducting detector that offers high energy resolution based on their sharp resistance-temperature feature in the superconductingto- normal transition. TES X-ray microcalorimeters have typically been designed and used for spectroscopic applications. In thiswork, we present a design optimization for a TES X-ray microcalorimeter array for high-energy scattering and diffraction measurements. In particular, Compton scattering provides information about the electron momentum distribution, while energy dispersive diffraction provides structural information about dense engineering materials. Compton scattering and energy dispersive diffraction experiments must be conducted in the very hard X-ray regime (similar to 100 keV), demanding a high X-ray stopping power in the detector; therefore, an absorber with a large heat capacity is needed in conjunction with the TES. In addition, both applications would benefit from an array composed of parallel strips. We present a design for a TES X-raymicrocalorimeter optimized for such applications. In particular, we model the longitudinal position dependence due to the finite thermal diffusion time in the absorber.
更多
查看译文
关键词
Compton scattering, energy dispersive diffraction, transition-edge sensors, position dependence