Portable Dc Voltage And Resistance Reference With Switching Unit For Calibration Of Electrical Precision Multifunction Instruments

2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018)(2018)

Cited 1|Views2
No score
Abstract
A portable temperature controlled DC Voltage and DC Resistance Reference with Switching Unit (RSU) has been developed at National Institute of Metrological Research (INRIM) to calibrate multifunction electrical instruments involving 10 V, a 1 Omega and a 10 k Omega Standards. The resistors are made with two resistors nets while the 10 V is a low noise-drift INRIM-projected circuit. Preliminary measurement over one-month period shows standards uncertainties span from 3.0x10(-7) to 6.4x10(-7), suitable for artifact calibration. With a real clock calendar, the RSU can show both the RSU Standards calibration values and updated ones between two calibrations according to an internal algorithm.
More
Translated text
Key words
DC Voltage, DC Resistance, calibrator, multimeter, artifact calibration, measurement uncertainties