Development of a 3-D Position Sensitive Neutron Detector Based on Organic Scintillators with Double Side SiPM Readout

2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC)(2018)

引用 2|浏览11
暂无评分
摘要
A 3-D position sensitive neutron detector is being developed based on a plastic scintillator array. A double side SiPM readout is used to determine the depth of interaction (DOI) in each scintillator unit. In the preliminary test, the DOI in a 254 x 6 x 6 mm3 SP101 plastic scintillator is measured at different positions using a collimated Co-60 source. The SiPM (KETEK PM6660) signals are recorded by a 2.5 GS/s digital oscilloscope. The DOI results are calculated using both the amplitude and the temporal information. Position resolutions (FWHM) of 2.5 cm and 6.6 cm are realized, respectively. A detector based on a 2-D array is capable of recording the 3-D position information of the incident neutron. The 3-D detector is to be used together with a neutron time projection chamber as a directional fast neutron detector. According to the simulation results, the angular resolution (8 degree FWHM) is much better than that of a typical neutron scatter camera.
更多
查看译文
关键词
neutron time projection chamber,directional fast neutron detector,3-D position sensitive neutron detector,organic scintillators,double side SiPM readout,plastic scintillator array,scintillator unit,3-D position information,incident neutron,SiPM signals,depth-of-interaction,SP101 plastic scintillator,typical neutron scatter camera,Co