Neuro control – using neuro monitoring data to control networked active instruments

Current Directions in Biomedical Engineering(2017)

引用 0|浏览12
暂无评分
摘要
AbstractIntraoperative neurophysiological monitoring (IONM) can be used to monitor the neural integrity and help the surgeon to reduce iatrogenic patient damage. We present a networked digital assistance system based on real-time IONM data. The system automatically reduces the power of active medical ablation instruments when they endanger neural structures. We integrated a nerve monitor into a real-time network of medical devices. In doing so, neurophysiological signals are available to all connected medical devices in real-time. With the IONM data a so-called function module Neuro Control calculates and controls the maximum allowed power for active instruments like an ultrasonic dissector.
更多
查看译文
关键词
neuro control,neuromonitoring,or.net,srtb,real-time network,medical devices,interoperability