Final Report on the APMPMT-S1 Supplementary Comparison Measurand Torque: 0 N center dot m, 1000 N center dot m, 2000 N center dot m (bilateral comparison).

Metrologia(2016)

引用 0|浏览1
暂无评分
摘要
The National Metrology Institute of Japan, AIST (NMIJ/AIST), Japan, and the Research Center for Metrology, LIPI (RCM-LIPI), Indonesia, had agreed to organize an inter-laboratory comparison program in the torque measurement at 1 kN·m and 2 kN·m for the torque measuring device. The objective of the project is to compare the performance of torque standards in participating institutes, in the torque range from 1 kN·m to 2 kN·m for torque measuring devices. The agreement between the results of two participants was very good. The smallest relative standard uncertainties of the machines stated by the participants are 6.6 × 10-5 and 3.0 × 10-4. The results of this comparison will be used supporting evidence for CMCs in this range. Main text To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).