Performances of different efficiency calibration methods of high-purity-germanium gamma-ray spectrometry in an inter-comparison exercise

Nuclear Science and Techniques(2019)

Cited 1|Views15
No score
Abstract
This study reports the performances of efficiency calibrations for high-purity-germanium gamma-ray spectrometry using the source-, Laboratory Sourceless Object Calibration Software (LabSOCS)- and ANGLE-based methods in an inter-comparison exercise. Although the results of LabSOCS and ANGLE for 241 Am emitting low-energy gamma rays were not very satisfactory, all of the three efficiency calibration methods passed acceptance criteria. The results confirmed the reliability of the calculation codes ANGLE and LabSOCS as alternative efficiency calibration methods in high-purity-germanium gamma spectrometry. This study is likely to promote the further application of the ANGLE and LabSOCS calculation codes in radioactivity measurements.