Chrome Extension
WeChat Mini Program
Use on ChatGLM

An extended core nanocoax pillar architecture for enhanced molecular detection.

L A D'Imperio, A E Valera, J R Naughton, M M Archibald,J M Merlo, T J Connolly,M J Burns,T C Chiles,M J Naughton

Biosensors and Bioelectronics(2019)

Cited 1|Views20
No score
Abstract
Biosensors that incorporate nanomaterials and nanofabrication techniques enable molecular detection of chemical and biological macromolecules with a high degree of specificity and ultrasensitivity. Here, we present a novel fabrication process that yields a nanostructure capable of detecting biological macromolecules. The extended core nanocoax (ECC) structure builds on a previously reported nanocoaxial-based sensor. The fabrication of the device incorporates an extended inner pillar, with controllable extension above the annulus and into the surrounding solution. This new design eliminates structural constraints inherent in the original nanocoax architecture. We also provide results demonstrating improvement in biosensing capability. Specifically, we show the capability of the new architecture to detect the B subunit of the Vibrio cholerae toxin at improved sensitivity (100 pg/ml) in comparison to optical enzyme-linked immunosorbant assay (1 ng/ml) and previously reported coaxial nanostructures (2 ng/ml).
More
Translated text
Key words
Nanocoax,Nanofabrication,Extended-core,Biosensor,High-sensitivity,ELISA
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined