Feasibility of automated fetal fractional shortening measurement with two-dimensional tracking and construction of a reference range for normal fetuses

Journal of medical ultrasonics (2001)(2019)

引用 6|浏览9
暂无评分
摘要
Purpose To assess the feasibility of an automated fractional shortening (Auto FS) measurement method using two-dimensional tracking and to construct a reference range for normal fetuses. Methods This study was conducted from May 2017 to March 2018. First, cardiac motion in the four-chamber view was recorded in the B-mode. Subsequently, the region of interest was set on the edge of the ventricular septum or ventricular muscle at a point one-third away from the atrioventricular valve toward the cardiac apex. Tracking was automatically performed. Values measured between the ventricular septum and right ventricle were defined as R-Auto FS, whereas those measured between the ventricular septum and left ventricle were defined as L-Auto FS. Those on each ventricular muscle were defined as Combined-Auto FS. Results Data were obtained from 131 singleton fetuses. R-Auto FS significantly decreased with an increase in the number of gestational weeks, and L-Auto FS and Combined-Auto FS showed the same tendency (Spearman’s correlation analysis: p = − .528, p = − .351, and p = − .636, respectively). Conclusion We succeeded in defining a reference Auto FS value for normal singleton pregnancies. Auto FS was negatively correlated with gestational age. This novel technique can assess fetal heart contractility.
更多
查看译文
关键词
Auto FS,Fractional shortening,Two-dimensional tracking method,Fetal cardiac function
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要