Calibration of high voltages at the ppm level by the difference of ^83m Kr conversion electron lines at the KATRIN experiment

M. Arenz, W.-J. Baek, M. Beck, A. Beglarian,J. Behrens, T. Bergmann, A. Berlev, U. Besserer,K. Blaum, T. Bode,B. Bornschein,L. Bornschein, T. Brunst,N. Buzinsky,S. Chilingaryan, W. Q. Choi, M. Deffert,P. J. Doe, O. Dragoun, G. Drexlin, S. Dyba, F. Edzards, K. Eitel,E. Ellinger, R. Engel, S. Enomoto,M. Erhard, D. Eversheim,M. Fedkevych,S. Fischer,J. A. Formaggio, F. M. Fränkle,G. B. Franklin, F. Friedel, A. Fulst,W. Gil, F. Glück,A. Gonzalez Ureña,S. Grohmann, R. Grössle, R. Gumbsheimer, M. Hackenjos,V. Hannen, F. Harms,N. Haußmann, F. Heizmann,K. Helbing,W. Herz, S. Hickford, D. Hilk, D. Hillesheimer, M. A. Howe,A. Huber, A. Jansen, J. Kellerer,N. Kernert, L. Kippenbrock, M. Kleesiek,M. Klein,A. Kopmann, M. Korzeczek,A. Kovalík, B. Krasch,M. Kraus, L. Kuckert, T. Lasserre,O. Lebeda, J. Letnev,A. Lokhov, M. Machatschek,A. Marsteller,E. L. Martin,S. Mertens,S. Mirz, B. Monreal,H. Neumann,S. Niemes, A. Off, A. Osipowicz,E. Otten, D. S. Parno,A. Pollithy, A. W. P. Poon,F. Priester, P. C.-O. Ranitzsch,O. Rest, R. G. H. Robertson, F. Roccati, C. Rodenbeck,M. Röllig, C. Röttele,M. Ryšavý, R. Sack,A. Saenz, L. Schimpf,K. Schlösser,M. Schlösser, K. Schönung, M. Schrank, H. Seitz-Moskaliuk, J. Sentkerestiová, V. Sibille,M. Slezák, M. Steidl,N. Steinbrink,M. Sturm, M. Suchopar, M. Suesser, H. H. Telle,L. A. Thorne, T. Thümmler, N. Titov,I. Tkachev,N. Trost,K. Valerius, D. Vénos, R. Vianden,A. P. Vizcaya Hernández,M. Weber,C. Weinheimer,C. Weiss, S. Welte, J. Wendel,J. F. Wilkerson,J. Wolf, S. Wüstling, S. Zadoroghny

European Physical Journal C(2018)

引用 23|浏览24
暂无评分
摘要
The neutrino mass experiment KATRIN requires a stability of 3 ppm for the retarding potential at − 18.6 kV of the main spectrometer. To monitor the stability, two custom-made ultra-precise high-voltage dividers were developed and built in cooperation with the German national metrology institute Physikalisch-Technische Bundesanstalt (PTB). Until now, regular absolute calibration of the voltage dividers required bringing the equipment to the specialised metrology laboratory. Here we present a new method based on measuring the energy difference of two ^83m Kr conversion electron lines with the KATRIN setup, which was demonstrated during KATRIN’s commissioning measurements in July 2017. The measured scale factor M=1972.449(10) of the high-voltage divider K35 is in agreement with the last PTB calibration 4 years ago. This result demonstrates the utility of the calibration method, as well as the long-term stability of the voltage divider.
更多
查看译文
关键词
conversion electron lines,high voltages,ppm level
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要