Nanoparticle-based lateral flow biosensor combined with multiple cross displacement amplification for rapid, visual and sensitive detection of Vibrio cholerae.

FEMS MICROBIOLOGY LETTERS(2017)

引用 7|浏览9
暂无评分
摘要
Vibrio cholerae is an important human pathogen that is responsible for cholera, a severe acute watery diarrhea. In the current study, a multiple cross displacement amplification (MCDA) coupled with amplicon detection by chromatographic lateral flow biosensor (LFB) method (MCDA-LFB) was successfully established and evaluated for the identification of V. cholerae. A set of 10 primers was designed specifically to recognize 10 different regions of the V. cholerae-specific gene ompW. The optimized time and temperature conditions for the MCDA were 30 min and 63 degrees C, respectively. The MCDA-LFB assay correctly identified 31 strains of V. cholerae but did not detect 13 non-cholerae Vibrio strains and 30 non-Vibrio strains. The sensitivity of MCDA-LFB for target pathogen detection in pure culture was 10 fg per reaction. In the case of spiked shrimp samples without enrichment, the limit of detection was 4.1 CFUs per reaction or equivalent to 4.1 x 10(2) CFU g(-1). The whole process, including shrimp homogenates processing (30 min), MCDA reaction (30 min) and results reporting (2 min), could be finished within 65 min. These results show that this assay is suitable for the rapid, sensitive and specific detection of V. cholerae in food, environmental and clinical samples.
更多
查看译文
关键词
Vibro cholerae,multiple cross displacement amplification,lateral flow biosensor,MCDA-LFB,limit of detection
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要